Unable to remap reference system

Question: 

The message "Unable to remap reference system" is displayed in the Comparison window. Why do I see this message instead of the gel lanes?

Answer: 

This message is produced when your gels are normalized using different reference systems, i.e. reference systems that have at least one different reference position, so that the reference systems can not be remapped automatically. In that case, you need to specify the molecular weight regression curves for both reference systems. Once the MW regression is known for both systems, the program will be able to remap the gels from one system into the other, and you will be able to compare the patterns in a Comparison window.

For defining regression curves, double-click on the fingerprint type, and then double-click on the reference system. If you entered the molecular weights for the names for the reference bands, you can simply choose Metrics > Copy markers from reference system, if not, you will have to enter the MW markers one by one.

Package(s): 
BIONUMERICS
Applicable for: 
Version 1.0 - 8.1